Feline NT-proBNP/Cardiac Troponin I Combo
Feline NT-proBNP/Feline Cardiac Troponin I Rapid Quantitative Test is a fluorescence immunoassay used along with Healvet 3000 Analyzer or quantitative determination of fNT-proBNP and fcTnI concentration in feline serum or plasma specimen. The test is used as an aid to myocardial function diagnosis and diagnosis of cardiomyocyte injury.
Specifications
| Animal | Feline |
|---|---|
| Sample type | Serum,plasma |
| Sample volume | 75μL |
| 39-86 °F storage | 24 months shelf life |
| the reference | check details in product description |
Reference range of fNT-proBNP in feline blood:
Detection range: 50-1500 pmol/L
Reference range:
Interpretation | pmol/L |
Low risk | <100 |
Suspicion of heart failure | 100-270 |
High risk heart failure | ≥270 |
Reference range of fcTnI in feline blood:
Detection range: 0.02-16.5 ng/mL
Reference range:
Interpretation | ng/mL |
Normal | <0.16 |
Suspicion of myocardial injury | 0.16-0.25 |
High risk of myocardial injury | >0.25 |
Operational Steps
Sample Preparation
Add 75 µL of serum or plasma sample into the buffer tube.
Mix thoroughly without forming bubbles.
Sample Application
Transfer 75 µL of the prepared mixture into the sample well of the test card.
Result Analysis
Insert the test card into the Healvet 3000 Analyzer with standard test mode.
Ready to implement Feline NT-proBNP/Cardiac Troponin I Combo?
Talk to our team about adding this test to your in-house diagnostic workflow.